OverviewThe MONET-R FT-NIR analyzer is a portable Fourier transform near-infrared spectrometer designed for rapid quantitative and qualitative analysis across laboratory and production environments. Covering 833-2597 nm with an InGaAs detector and patented interferometer, it delivers reproducible spectra suitable for multicomponent testing and process control. Integrated software supports instrument control, one-click multi-parameter testing and model development for routine industrial use.
Product Features- Full spectral coverage 833–2597 nm for key chemical bond information (N-H, S-H, C-H, C=O)
- Fast analysis: simultaneous multi-parameter detection in under 1 minute (adjustable)
- User-oriented software combining instrument control, data processing and modeling
- Integrating-sphere diffuse-reflection sampling with rotatable/translatable stage for representative spectra
- Long-life light source and pre-aligned consumables for quick field maintenance
- Built-in reference material for automatic calibration and wavelength monitoring
- Multiple sample cups and accessories to handle powders, liquids and solids
Parameter Standards- Wavelength range: 833-2597 nm
- Instrument principle: Fourier Transform (FT)
- Detector: InGaAs detector
- Spectral resolution: 4 cm⁻¹
- Wavenumber accuracy: < 0.05 cm⁻¹
- Wavenumber repeatability: < 0.006 cm⁻¹
- Absorbance accuracy: < 0.1 %T
- Scanning speed: 1 minute (adjustable)
- Protection rating: IP54
Industry Applications- Feed and animal nutrition analysis
- Grain, oilseed and milling quality control
- Breeding and seed evaluation
- Chemical process quality monitoring
- Laboratory and on-site industrial sample analysis
Development Reference Standards- ASTM E1655-05 — Infrared multivariate quantitative analysis
- ASTM E168-06 — Techniques for infrared quantitative analysis
- ASTM E1790-04 — Near infrared qualitative analysis
- 39-00 — Near-Infrared Methods: Guidelines for model development and maintenance
- 39-10 — NIR method for protein in small grains
- 39-11 — NIR method for protein in wheat flour
- 39-20 — NIR method for protein and oil in soybeans
- 39-21 — NIR method for whole-grain soybean analysis
- 39-25 — NIR method for protein in whole-grain wheat
- 39-70A — NIR method for hardness in wheat
- ICC-159 — Determination of protein by NIR spectroscopy
Technical Specifications- Wavelength range: 833-2597 nm
- Instrument principle: Fourier Transform (FT)
- Detector: InGaAs detector
- Spectral resolution: 4 cm⁻¹
- Wavenumber accuracy: < 0.05 cm⁻¹
- Wavenumber repeatability: < 0.006 cm⁻¹
- Absorbance accuracy: < 0.1 %T
- Scanning speed: 1 minute (adjustable)
- Protection rating: IP54
- Dimensions: L 405 mm × W 330 mm × H 290 mm
- Weight: 30 kg
- Power supply: AC 220V 50Hz
- Operating conditions: 5–35°C / <80% RH
- Warm-up time: none, ready at startup