OverviewThe IAS-7500 is a laboratory near-infrared (NIR) analyzer designed for rapid quantitative and qualitative analysis of granular, powdered and liquid samples. Its compact footprint and large-area scanning mode improve sampling representativeness and spectrum quality, enabling reliable real-time monitoring of sample variability for laboratory benchmark measurements.
Key features- Fast analysis: multi-indicator results within 1 minute; batch processing suitable for high sample throughput and limited staff.
- Multi-form sample capability: single analyzer supports granular, powder and liquid matrices.
- Large-area scanning mode: enhances representativeness of sampling and improves spectral quality.
- High precision and repeatability for laboratory-standard measurements.
ApplicationsFeed industry: detection of moisture, protein, fat, fiber (ADF, NDF), protein solubility, starch, acid value and freshness (VBN) in raw materials and finished feed, including pet food and forage.
Grain & oil industry: analysis of protein, fat, moisture, fiber, ash, acid value and peroxide value in cereals, oilseeds, edible oils and by-products.
Flour industry: measurement of moisture, ash, protein, gluten and water absorption in flour.
Chemical industry: organic composition analysis, catalyst development and reaction monitoring, non‑destructive polymer characterization (molecular weight distribution, branching degree, end-group concentration), oil and fuel analysis.
Sampling accessory7500-Medium Sample Cell | Material: Aluminum alloy (6061). Dimensions: D: 138 mm, H: 25 mm (excluding handle). Use: small granular samples (e.g., feed pellets) and powders (e.g., flour, rice bran).
7500-Transflection Cover | Material: Aluminum alloy (6061). Dimensions: D: 38 mm, H: 8 mm (excluding handle). Optical path: 1 mm. Use: transflection cover for small cells; suitable for transparent liquids.
7500-Small Cell Holder | Material: Aluminum alloy (6061). Dimensions: D: 136 mm, H: 8.4 mm. Use: holder plate for small sample cells.
7500-Large Sample Cell | Material: Aluminum alloy (6061). Dimensions: D: 138 mm, H: 50 mm. Use: powders, granular or grass-type samples.
7500-Large Cell Press | Material: Stainless steel. Dimensions: D: 128 mm, H: 15 mm (excluding handle). Use: press for testing grass-type samples with the large cell.
Reference standards- ASTM E1655-05 — Standard Practices for Infrared Multivariate Quantitative Analysis
- ASTM E168-06 — Standard Practices for General Techniques of Infrared Quantitative Analysis
- ASTM E1790-04 — Standard Practice for Near Infrared Qualitative Analysis
- 39-00 — Near-Infrared Methods: Guidelines for Model Development and Maintenance
- 39-10 — NIR Reflectance Method for Protein Determination in Small Grains
- 39-11 — NIR Reflectance Method for Protein Determination in Wheat Flour
- 39-20 — NIR Reflectance Method for Protein and Oil Determination in Soybeans
- 39-21 — NIR Reflectance Method for Whole-Grain Analysis in Soybeans
- 39-25 — NIR Reflectance Method for Protein Content in Whole-Grain Wheat
- 39-70A — NIR Reflectance Method for Hardness Determination in Wheat
- ICC-159 — Determination of Protein by Near Infrared Reflectance (NIR) Spectroscopy
Technical specifications- Wavelength range (NIR): 920–1660 nm
- Detector: InGaAs (Indium Gallium Arsenide)
- Spectroscopic principle: grating + array detector
- Baseline noise: ≤ 0.0001 AU
- Wavelength repeatability: ≤ 0.005 nm
- Wavelength accuracy: ≤ 1 nm
- Absorbance repeatability: ≤ 0.0001 AU
- Measurement time: ≤ 1 min
- Data interval: 1 nm
- Dimensions (L×W×H): 388 mm × 360 mm × 274 mm
- Weight: 15 kg
- Light source: tungsten halogen lamp
- Power supply: 24 V / 3 A
- Power consumption: 24 V / 2.5 A
- Display: 7-inch touch screen
- Operating system: built-in Linux
- Operating temperature: 5–45 °C
- Storage temperature: -20–60 °C
- Operating humidity: <93% RH