OverviewRA-915F provides accurate, reliable mercury (Hg) measurements by direct thermal decomposition across a broad range of applications. Complex-matrix samples can be analyzed as-is without chemical pre-treatment, reducing time and consumable costs.
Principle of operationDirect thermal decomposition followed by atomic absorption spectrometry at 254 nm with Zeeman background correction ensures selective, sensitive, interference-free measurements without matrix effects.
Features and benefits- Fast direct analysis of solids, liquids and biosamples
- Low detection limit and high selectivity due to Zeeman background correction
- Real-time feedback system for sample decomposition control extends dynamic range and eliminates background interferences
- Broad measurement range: more than 5 orders of magnitude
- No sample pretreatment or reagents required, lowering analysis cost
- Safe and simple operation: no compressed gases required (air used as carrier gas)
- No pre-concentration on gold traps, increasing analysis speed and reducing operational complexity
- No memory effect: low-concentration samples can be analyzed immediately after high-concentration samples
- Programmable furnace for optimized sample decomposition and mercury thermospeciation studies
- Compact footprint: suitable for mobile laboratories
ApplicationsRA-915F is used for direct determination of mercury concentration in:
- Soils, sediments, rocks, ores, minerals
- Food, nutritional additives, animal feed
- Crude oil, oil products, naphtha, petrochemicals
- Biosamples: tissues, hair, urine, blood, etc.
- Wastewaters, wastes, sludge
- Coal, coke, fly ash, gypsum
- Sorbent traps for stack gases and passive air samplers
- Drugs, cosmetics, plastics, fertilizers and many other materials
StandardsThe direct thermal decomposition technique complies with standards including ASTM D6722, D7622, EN/TS 17286, US EPA 7473, 30B, PS12B, SN/T 4429.2-2016, HJ 923-2017 and others.
SoftwareRAPID is the dedicated software for RA-915 analyzers. RAPID complies with FDA 21 CFR Part 11 requirements and provides:
- Data acquisition, processing and reporting
- Real-time visualization of selective and non-selective absorbance
- Databases for calibrations, processing and storage of measurement data
- Adjustable modes for optimal thermal decomposition of specific matrices
WarrantyRA-915F is covered by a full 1-year warranty.
ServicesInstallation and commissioning are available on request by service engineers or certified distributors. Support and operator training are provided by application specialists to meet customer requirements.
Technical specifications- Measurement principle: Direct thermal decomposition followed by atomic absorption spectrometry at 254 nm with Zeeman background correction
- Detection limit: 0.5 ppb (µg/kg)
- Measurement range: 0 – 200,000 ppb (µg/kg)
- Sample boat capacity: 1000/2000 mg (solids) and 500/1000 µl (liquids)
- Analysis time: 1 – 5 min
- Analyzer warm-up time, max: 40 min
- Temperature range of atomizer and analytical cell: 150 °C – 800 °C
- Power supply: 100 – 240 VAC, 50 – 60 Hz
- Power consumption, max: 700 W
- Dimensions (WxDxH): spectrometer unit 175×110×235 mm; thermal unit 430×340×135 mm; power supply unit 400×280×135 mm
- Weight: 21.5 kg
- Detection limits by matrix:
- Soils, rocks, sediments: 0.5 ppb (µg/kg)
- Water and solutions: 0.2 ppb (µg/L)
- Foodstuff: 0.5 ppb (µg/kg)
- Biological samples (tissues, blood, hair, etc.): 0.5 ppb (µg/kg)
- Crude oil, gas condensate: 1 ppb (µg/kg)
- Naphtha direct: 1 ppb (µg/kg); naphtha preconcentration: 0.05 ppb (µg/kg)
- Coal: 1 ppb (µg/kg)
- Sorbent traps for stack gases: 0.5 ng* (range 0.5 - 50,000 ng)